R&S PR100 User Manual
[SENSe]
. :MSCan
. . :HOLD
. . . :TIME <numeric_value>|MINimum|MAXimum
Setting the hold time for a signal-controlled scan continuation. If the signal
disappears during the dwell time, the hold time is started. After completion of
the hold time, the scan is continued with the next frequency even if the dwell
time has not yet been completed. If the signal exceeds the squelch threshold
during the hold time, the hold time is reset and the end of the dwell time or the
renewed disappearance of the signal is awaited. The hold time is only used if
the control function "STOP:SIGNal" (see SENSe:MSCan:CONTrol) is
switched on.
Setting the time to 0 (zero) has the same effect as switching off the control
function with SENS:MSC:CONT:OFF “STOP:SIGN”.
This value is mapped onto the same variable as the frequency scan hold time
variable, meaning that a change affects both scan modes.
Parameters:
<numeric_value> hold time in seconds
MINimum|MAXimum lowest|highest hold time
Example:
SWEep:HOLD:TIME 100 ms
[SENSe]
. :MSCan
. . :HOLD
. . . :TIME? [MINimum|MAXimum]
Query of hold time. This command is an alias of SENS:SWE:HOLD:TIME?.
Parameters:
none query of current hold time
MINimum|MAXimum query of lowest|highest hold time
Result:
Hold time in seconds
Example:
MSCan:HOLD:TIME? 0.10