28 Cubis
®
MCE Operating Instructions
System Settings
7 System Settings
7.1 Performing System Settings
Default settings can be adjusted for the device and the applications in order to align
with the ambient conditions and individual operating requirements.
The following settings are necessary to operate the device together with connected
components:
− Set up the communication of the connected devices
− Set up additional components
The following settings are recommended to set up the device:
− Set the menu language
− Set the behavior of the isoCAL function
− Set the behavior of the motorized draft shield (only for devices with a motorized
draft shield)
Procedure
t Press the [Menu] button.
t Open the desired settings menu.
t To adjust settings: Open the desired main menu and submenu.
t Select and confirm the desired configuration value (configuration values, see
Chapter “7.3 Parameter List”, page 29).
t Exit the menu.
y If settings have been adjusted in the “SETUP”, “DEVICE”, “DATA.OUTP.” or
“APPLIC.” menu: The [BOOTING] display appears in the operating display and the
device restarts.
7.2 Switching Off the isoCAL Function
If the isoCAL function is switched off for a conformity-assessed device, the device
can only be used for legal-for-trade applications in restricted temperature ranges
(see Chapter “15.6.2 Ambient Temperature for the isoCAL Function”, page 58). The
isoCAL function cannot be switched off for all model versions.
Procedure
t In the “SETUP” / “BALANCE” menu, for the “ISOCAL” parameter, select the “OFF”
configuration value.