62 Cubis
®
 MCE Operating Instructions
  Technical Data
15.8  Recommended Calibration Weight
MCE14202S  MCE14202P MCE10202S MCE8202S
Unit Value Value Value Value
External test weight g 14000 14000 10000 7000
Recommended accuracy class E2 E2 E2 E2
MCE6202S  MCE6202P MCE4202S MCE2202S
Unit Value Value Value Value
External test weight g 5000 5000 3000 1500
Recommended accuracy class E2 E2 E2 E2
MCE1201S  MCE12201S MCE8201S MCE5201S
Unit Value Value Value Value
External test weight g 700 12000 8000 5000
Recommended accuracy class E2 F1 F1 F1
15.9  isoCAL Function
15.9.1  Models MCE14202S | MCE14202P | MCE10202S
Unit Value 
isoCAL is triggered by the following criteria: 
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling
15.9.2  Models MCE8202S | MCE6202S | MCE6202P | MCE4202S | MCE2202S | MCE1202S
Unit Value 
isoCAL is triggered by the following criteria: 
In the event of a temperature change K 2
After a time interval h 12
After successful leveling
15.9.3  Models MCE12201S | MCE8201S | MCE5201S
Unit Value 
isoCAL is triggered by the following criteria: 
In the event of a temperature change K 4
After a time interval h 12
After successful leveling