Cubis
®
MCE Operating Instructions 31
System Settings
Parameters Setting values Explanation
CAL./ADJ. EXT.CAL. The [Adjust] button starts an external calibration with the preset calibration weight.
E.CAL.USR. The [Adjust] button starts an external calibration with the user-defined calibration
weight value.
INT.CAL.* The [Adjust] button starts an internal calibration.
CAL.SEQ. ADJUST* Calibration and adjustment is one routine.
CAL.-ADJ. Adjustment must be started or exited manually after calibration with the [Adjust]
button.
ON Z/T ON* Activates the initial taring / zeroing. The device is tared or zeroed after it is switched on.
OFF Deactivates the initial taring / zeroing: After it is switched on, the device shows the value
before it was last switched off.
ISOCAL OFF Switches the isoCAL function off.
NOTE TO The [isoCAL] button flashes if the balance needs to be adjusted. The isoCAL function
must be manually triggered with the [Adjust] button.
ON* Activates the isoCAL function. The device is automatically adjusted as soon as a trigger
starts the isoCAL function.
CAL.UNIT GRAMS* Changes the calibration weight unit to grams.
KILOGR. Changes the calibration weight unit to kilograms.
* Factory setting
Parameters in the “General Services” Submenu
Parameters Setting values Explanation
MEN.RESET YES Resets the system settings to the factory default settings.
NO* Deactivates the option of resetting the device menu.
* Factory setting
7.3.2 Parameters in the “Device” Main Menu
Parameters in the “Extras” Submenu
Parameters Setting values Explanation
MENU EDITABLE* Activates write access: The menu parameters can be changed.
RD.ONLY Activates read access: The menu parameters cannot be changed.
SIGNAL OFF Switches the acoustic signal off.
QUIET Sets the volume of the acoustic signal to “quiet”.
MEDIUM* Sets the volume of the acoustic signal to “medium”.
LOUD Sets the volume of the acoustic signal to “loud”.
KEYS UNLOCKED* Deactivates the key lock.
LOCKED Activates the key lock.
ON MODE ON/STDBY* The [On / Off] key switches between on / standby.
ON/OFF The [On / Off] key switches between on / off.
AUTO ON Changes the function of the [On / Off] key: The device no longer switches off or to
standby mode, instead it starts a boot process.
LIGHTING OFF Deactivates the lighting on the operating display.
ON* Activates the lighting on the operating display.
* Factory setting