42 Cubis
®
MCE Operating Instructions
Operation
Possible conditions are:
− The ambient temperature has changed since the last adjustment / calibration.
− The interval time was exceeded (interval time, see Chapter “15.8 Recommended
Calibration Weight”, page 62).
− The device has been leveled.
− The device has been disconnected from the power supply since the last
adjustment / calibration (only for conformity-assessed models).
Procedure
t If the isoCAL auto-start function is set (configuration values for the “ISOCAL”
parameter, see Chapter “7.3.1 Parameters in the “Setup” Main Menu”, page 29):
y If the isoCAL function is triggered according to the aforementioned
conditions: The [isoCAL] button flashes in the operating display.
y In the operating display, a time display counts down from 15 seconds to 0.
y If no load change or no operation takes place on the device before the
expiration of the time display: The isoCAL function starts.
t If the isoCAL manual start function is set (configuration values for the “ISOCAL”
parameter, see Chapter “7.3.1 Parameters in the “Setup” Main Menu”, page 29):
y If the isoCAL function is triggered: The [isoCAL] button flashes in the
operating display.
t Press the [isoCAL] button.
y The isoCAL function starts.
8.7 Internally Calibrating and Adjusting the Device
Calibration and Subsequent Adjustment
Requirements
− The weighing pan is unloaded.
− The operating display shows a stable weight value.
Procedure
t To set the internal adjustment in the menu: In the “SETUP” / “BALANCE” menu,
for the “CAL./ADJ.” parameter, select the “INT.CAL.” configuration value.
t To set the sequence of the adjustment function to calibration and subsequent
automatic adjustment: In the “SETUP” / “BALANCE” menu, for the “CAL.SEQ.”
parameter, select the “ADJUST” configuration value.
t Exit the menu.
t Press the [Zero] key.
t To start the internal calibration / adjustment: Press the [Adjust] button.
t If automatic leveling is set for a model with motorized adjusting feet: The device
levels itself automatically.
y The internal calibration weight is loaded.
y The [
CAL.RUN.
] display appears in the operating display during calibration.
y The device is automatically calibrated and the internal calibration weight is
removed.
y The [CAL.END] display indicates the end of calibration.