MCE Operating Instructions
Operation
8.11 Running Applications (Examples)
8.11.1 Executing the “Toggle Between Weight Units” Function
The “Toggle between weight units” function enables a switch between the basic unit
display and up to 4 other units. The units can be set during the weighing process and
the decimal places can be adjusted.
The basic unit is displayed every time the device is started.
Procedure
t To activate the “Toggle between weight units” function in the menu: In the
“BALANCE” menu, for the “UNIT” parameter, select the “ON” configuration value.
t Exit the settings menu.
y The device boots up.
t To select the convertible units: Press and hold down the “Toggle between weight
units” button of the “Change” key.
t The “Toggle between weight units” function menu opens.
t Select a submenu from “Unit 1” – “Unit 4”.
t Select the desired unit and confirm the display (available units, see
Chapter “7.3.7 Parameters in the “Unit 1”, “Unit 2”, “Unit 3”, and “Unit 4”
Function Menus”, page 38).
t To set the number of display digits for the selected unit: Press the “Confirm”
button (configuration parameters, see Chapter “7.3.7 Parameters in the “Unit 1”,
“Unit 2”, “Unit 3”, and “Unit 4” Function Menus”, page 38).
t Select the desired number of display digits and confirm the selection.
t Exit the “Toggle between weight units” function menu.
y The current weight value is displayed in the selected unit.
t To switch the displayed mass unit during weighing or before starting an
application: Press the “Toggle between weight units” button or the “Change” key
multiple times if necessary, until the desired unit is displayed.
y The current weight value is displayed in the selected unit.
8.11.2 Running the “Statistics” Application
The “Statistics” application saves up to 99 weight values and evaluates these
statistically.
The following values are saved and exported by the statistics application:
Number of components
Mean value
Standard deviation
Variation coefficient
Sum of all values
Lowest value (minimum)
Highest value (maximum)
Spread: Difference between maximum and minimum
The “Statistics” application can be combined with the following functions:
Toggle between weight units only available before starting the application
ID marking
Automatic taring

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCE Series and is the answer not in the manual?

Sartorius Cubis MCE Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCE Series
CategoryScales
LanguageEnglish

Related product manuals