Cubis
®
 MCE Operating Instructions  47
  Operation 
Procedure
 t Select the Statistics application in the “APPLIC.” menu.
 t Exit the settings menu.
 t Place a sample on the weighing pan.
 t Change the displayed weight unit if necessary.
 t To start the recording of the statistics: Press the [Confirm] button on the 
keypad.
 y The current weight value is saved.
 t To save the next value: 
 t Remove the sample on the weighing pan. 
 t Place a new sample on the weighing pan and press the [Confirm] button.
 t To switch between the display of the current weight value, the number of saved 
components, and the calculated mean in the results display: Press the [Up] or 
[Down] button.
 t To print and exit the current statistics, and to delete the saved values: Press the 
[Back] or [GLP] button.