Cubis
®
MCE Operating Instructions 47
Operation
Procedure
t Select the Statistics application in the “APPLIC.” menu.
t Exit the settings menu.
t Place a sample on the weighing pan.
t Change the displayed weight unit if necessary.
t To start the recording of the statistics: Press the [Confirm] button on the
keypad.
y The current weight value is saved.
t To save the next value:
t Remove the sample on the weighing pan.
t Place a new sample on the weighing pan and press the [Confirm] button.
t To switch between the display of the current weight value, the number of saved
components, and the calculated mean in the results display: Press the [Up] or
[Down] button.
t To print and exit the current statistics, and to delete the saved values: Press the
[Back] or [GLP] button.