58 Cubis
®
MCE Operating Instructions
Technical Data
15.6 Ambient Conditions
15.6.1 Installation Site
Unit Value
Installation site
Standard laboratory rooms
Installation site according to IEC 60259-1,
maximum altitude above sea level
m 3000
For indoor use only
Temperature
In operation °C +5 – +40
In operation for conformity-assessed devices:
See information on the device’s ID plate
During storage and transport °C -20 – +60
Relative humidity
At temperatures of up to 31 °C % 80
Then linear decrease from 80 % at 31 °C to 50 % at 40 °C
No heat from heating systems or direct sunlight
No drafts from open windows, AC systems, or doors
No vibrations
No “heavy traffic” areas (personnel)
No electromagnetic fields
No dry air
15.6.2 Ambient Temperature for the isoCAL Function
MCE14202S MCE14202P MCE10202S MCE8202S
Unit Value Value Value Value
Scope of application as per Directive 2014/31/EU
With isoCAL function °C +10 – +30
+10 – +30 +10 – +30
+10 – +30
Without isoCAL function °C +17 – +27
+17 – +27 +17 – +27
+10 – +30
MCE6202S MCE6202P MCE4202S MCE2202S
Unit Value Value Value Value
Scope of application as per Directive 2014/31/EU
With isoCAL function °C +10 – +30 +10 – +30 +10 – +30 +10 – +30
Without isoCAL function °C +10 – +30 +10 – +30 +10 – +30 +10 – +30
MCE1202S MCE12201S MCE8201S MCE5201S
Unit Value Value Value Value
Scope of application as per Directive 2014/31/EU
With isoCAL function °C +10 – +30 +10 – +30 +10 – +30 +10 – +30
Without isoCAL function °C +10 – +30 +10 – +30 +10 – +30 +10 – +30