34 Cubis
®
MCE Operating Instructions
System Settings
Parameters Setting values Explanations
STOPBIT 1 BIT* Sets the number of stop bits to 1.
2 BITS Sets the number of stop bits to 2.
HANDSHK. SOFTWARE Sets the handshake protocol to software handshake.
HARDWARE Sets the handshake protocol to hardware handshake.
NONE* Does not set a handshake protocol.
DATABIT 7 BITS Sets the number of data bits to 7.
8 BITS* Sets the number of data bits to 8.
* Factory setting
Parameters in the “USB” Submenu
Parameters Setting values Explanation
DEV.USED NONE* Indicates that no device connection is detected at the USB port.
[Device Designation] Displays the designation of the devices connected to the USB port.
* Factory setting
7.3.3 Parameters in the “Data Output” Main Menu
Parameters in the “SBI Communication” Submenu
Parameters Setting values Explanations
COM. OUTP. IND.NO* Starts the data output after pressing the key or software command as an individual
value without stability.
IND.AFTER Starts the data output after pressing the key or software command as an individual
value without stability.
IND.AT Starts the data output after pressing the key or software command as an individual
value at stability.
AUTO W/O Activates the automatic data output without stability.
AUTO W/ Activates the automatic data output after stability.
STOP.AUTO OFF* Deactivates the option to abort the automatic data output.
ON The automatic data output is aborted by pressing the [PRINT] key or a software
command.
AUTO.CYCL. EACH VAL* Starts the automatic data output with a cycle after each value.
2. VALUE Starts the automatic data output with a cycle after every 2nd value.
FORMAT 16 CHARS 16 characters per line are data output.
22 CHARS* Data output provides 22 characters per line (16 characters for the measured values
and 6 characters for identifiers).
EXTR.LINE Data output provides an additional line with the date, time, and weight value.
AUTO.TARE OFF* Deactivates automatic taring after data output.
ON The device automatically tares after data output.
* Factory setting