MCE Operating Instructions 17
Operating Concept
4.5 Menu Structure
4.5.1 Overview of the Device Settings Menu
t Navigating in Menus (see Chapter 4.6, page 21).
Level 1 Level 2 Level 3 Description
SETUP BALANCE AMBIENT
“Ambient conditions”
Enter the ambient conditions at the device’s installation
site.
APP FILT.
“Application filter”
Equalize the load deviations in the display. The load
deviations occur between the complete placement of the
sample on the weighing pan and reaching the final result
display.
STABIL.
“Stability range”
Define the accuracy of the stability range. If the weight
unit symbol appears, the weight readout is stable within
the range indicated by the stability range
ST.DEL.
“Stability delay”
Define the duration of the stability delay. The stability
delay is used to bridge residual fluctuations in the vibration
of the device in order to display reliable results.
ZERO/TAR.
“Zeroing and taring”
Define the trigger for zeroing and taring.
AUTOZER.
“Auto zero”
Activate / deactivate automatic control of the zero display.
The deviations of a set fraction of scale intervals per
second starting from the display zero point are
automatically zeroed.
UNIT Define the weight unit, e.g. grams, kilograms, pounds.
Depending on the country-specific model version, not all
weight units listed may be available. Some weight units
may be blocked from use in conformity-assessed devices,
depending on national verification laws.
DISP.DIG. Define whether all decimal places are displayed. A lower
display accuracy enables a faster display. The setting option
is not available on conformity-assessed devices.
CAL./ADJ.
“Adjust”
Define the function of the [Adjust] button.
CAL.SEQ.
“Calibration sequence”
Define the sequence of the “Calibrate and adjust” function.
ON Z/T
Tare / zero at power on
Activate / deactivate automatic taring when starting the
device.
ISOCAL Configure the isoCAL function.
CAL.UNIT
“Calibration weight unit”
Define the weight unit for the calibration weight.
GEN.SERV.
“General services”
MEN.RESET
“Menu reset”
Reset the menu to factory settings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCE Series and is the answer not in the manual?

Sartorius Cubis MCE Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCE Series
CategoryScales
LanguageEnglish

Related product manuals