20 Entris
®
II Advanced Line Operating Instructions
Operating Concept
4.11.2 Parameters in the “Settings / Date and Time” Menu
Parameters Settings Explanation
Date Defines the date.
Date format DD-MMM-YY Sets the date display format to DD.MMM.YY
MMM-DD-YY Sets the date display format to MMM.DD.YY
YY-MM-DD
(ISO)*
Sets the date display format to ISO FORMAT YYYY-MM-DD.
Sets the time to 24-hour mode.
Time Sets the time.
Time format** 24h* Sets the time to 24-hour mode.
12h (AM/PM) Sets the time to 12-hour mode.
* Factory setting
** Only active if “YY-MM-DD (ISO)” is not selected under “Date format”.
4.11.3 Parameters in the “Settings / Cal./Adj. Settings” Menu
Parameters Settings Explanation
isoCAL Off Switches the isoCAL function off.
Info, manual
start
If the balance needs to be calibrated and adjusted: The [isoCAL] button is
highlighted orange. The isoCAL function must be manually triggered with the
[isoCAL] button.
Automatic* Activates the automatic calibration and adjustment function as soon as
predefined time intervals or temperature values are exceeded.
Calibration
report
Displays max. 99 calibration reports per calendar day.
Displays max. 30 calendar days.
* Factory setting