42 Entris
®
 II Advanced Line Operating Instructions
 Malfunctions
10  Malfunctions
10.1  Warning Messages
Warning message Fault Cause Remedy Chapter, 
page
APP.ERR. The device has 
measured an invalid 
weight value.
The applied weight is too 
low.
Increase the applied weight 
to more than the minimum 
load.
The weight value is 
negative. 
No sample has been 
placed on the balance.
Place the sample on the 
balance.
DIS.ERR. The value to be 
output cannot be 
shown in the 
operating display.
The data to be displayed 
is not compatible with the 
set display format.
Adjust the display settings, 
e.g., resolution, unit, 
decimal places.
 The device is 
overloaded.
The device’s maximum 
weighing capacity has 
been exceeded.
Reduce the applied weight 
to below the device’s 
maximum weighing 
capacity.
 The modulation of the 
weighing converter 
inside the device is 
too low.
No weighing pan has 
been placed on the 
balance.  
Insert the weighing pan into 
the device and switch the 
device off and on again.
A previously forgotten 
weight was removed after 
starting the device.
An error exists in the 
weighing system or in the 
device electronics.
Contact Sartorius Service.
COMM.ERR. The device is not 
receiving any weight 
values.
No communication exists 
between the control unit 
and the weigh cell.
Wait until the control unit 
restores the communication 
with the weigh cell.
If the problem occurs again: 
Contact Sartorius Service.
PRT.ERR. The [Print] button is 
locked.
The data interface for 
print output is set to xBPI 
mode.
Reset the menu to the 
factory settings.
If the problem occurs again: 
Contact Sartorius Service.
SYS.ERR. The system data is 
faulty.
A memory error exists in 
the control unit.
Switch the device off and on 
again. 
8.1, 38
If the problem occurs again: 
Contact Sartorius Service.