24 Entris
®
 II Advanced Line Operating Instructions
  Operating Concept
Parameters Settings Explanation
Sample ID 
function
On   Activates the sample ID.
  The sample ID is queried before each printout.
Autom. 
increment
Counts up the sample ID automatically.
Autom. 
decrement
Counts down the sample ID automatically.
Off* Deactivates the sample ID.
Sample ID2 
function
On   Activates the sample ID2.
  The sample ID is queried before each printout.
Autom. 
increment
Counts up the sample ID2 automatically.
Autom. 
decrement
Counts down the sample ID2 automatically.
Off* Deactivates the sample ID2.
* Factory setting
4.11.10  Parameters in the “Settings / SBI Protocol” Menu
Parameters Settings Explanation
SBI function Manual 
without 
stability
If a print command is given, e.g., via the [print] button: The weight value is 
printed.
Manual with 
stability*
If a print command is given, e.g., via the [print] button: The weight value not is 
printed until the device is stable.
Auto without 
stability
Prints the weight value automatically without device stability.
Auto without 
stability
Prints the weight value automatically only when the device is stable.
Print after 
weight 
change
Prints after a threshold for device stability has been exceeded and the weight has 
previously been reduced to less than half of the threshold.
Manual print 
format
Value w/o 
identifier
Prints the weight value without labeling, e.g., “Net” or “G”.
Value* Prints the weight value.
Date, Value Prints the weight value and a time stamp.
Value (N, T, 
G#)
Prints a weight value block (net, tare, calculated gross).
Date, Value 
(N, T, G#)
Prints a time stamp and a weight value block (net, tare, calculated gross value).
* Factory setting