Entris
®
 II Advanced Line Operating Instructions  21
    Operating Concept 
4.11.4  Parameters in the “Settings / Weighing” Menu
Parameters Settings Explanation
Safety Level High* If the balance is not correctly leveled or a calibration/adjustment is not needed 
or the minimum initial weight requirements according to USP are not met: 
  A warning message is displayed. 
  No data is transmitted.
  The printout is blocked.
  Starting and saving in applications is blocked.
Standard If the balance is not correctly leveled or a calibration/adjustment is not needed 
or the minimum initial weight requirements according to USP are not met: 
  A warning message is displayed. 
Low   If the balance is not leveled: The [Start leveling] button is active. 
  If the balance needs to be calibrated and adjusted: The [isoCAL] button is 
active.
Ambient 
conditions
Very stable Sets the ambient conditions to “very stable”. Activates a fast change in the 
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
  Very stable table near the wall
  Closed and calm room
Stable* Sets the ambient conditions to “stable”. Recommended for the following work 
environment:
  Stable table
  Slight movement in the room
  Slight draft
Unstable Sets the ambient conditions to “unstable”: Activates the delayed change in 
weight values with a reduced output rate. Recommended for the following work 
environment: 
  Simple office desk
  Room with moving machinery or personnel
  Slight air movement
Very unstable Sets the ambient conditions to “very unstable”: Activates a significantly delayed 
change in the weight values and long wait for stability with a further reduction in 
the output rate.
Recommended for the following work environment:
  Noticeable and slow floor vibrations
  Noticeable building vibrations
  Weighed goods moved
  Very strong air movements
Application Weighing* Activates a filter that enables a fast change in the display for very fast load 
changes. Display changes with minimal load changes (in the digit range) occur 
more slowly.
Dosing Activates a filter that enables a very fast change in the display with minimal load 
changes, e.g., when dosing or filling vessels.
Stability signal High accuracy
Medium 
accuracy*
Fast
* Factory setting