Entris
®
II Advanced Line Operating Instructions 21
Operating Concept
4.11.4 Parameters in the “Settings / Weighing” Menu
Parameters Settings Explanation
Safety Level High* If the balance is not correctly leveled or a calibration/adjustment is not needed
or the minimum initial weight requirements according to USP are not met:
A warning message is displayed.
No data is transmitted.
The printout is blocked.
Starting and saving in applications is blocked.
Standard If the balance is not correctly leveled or a calibration/adjustment is not needed
or the minimum initial weight requirements according to USP are not met:
A warning message is displayed.
Low If the balance is not leveled: The [Start leveling] button is active.
If the balance needs to be calibrated and adjusted: The [isoCAL] button is
active.
Ambient
conditions
Very stable Sets the ambient conditions to “very stable”. Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
Very stable table near the wall
Closed and calm room
Stable* Sets the ambient conditions to “stable”. Recommended for the following work
environment:
Stable table
Slight movement in the room
Slight draft
Unstable Sets the ambient conditions to “unstable”: Activates the delayed change in
weight values with a reduced output rate. Recommended for the following work
environment:
Simple office desk
Room with moving machinery or personnel
Slight air movement
Very unstable Sets the ambient conditions to “very unstable”: Activates a significantly delayed
change in the weight values and long wait for stability with a further reduction in
the output rate.
Recommended for the following work environment:
Noticeable and slow floor vibrations
Noticeable building vibrations
Weighed goods moved
Very strong air movements
Application Weighing* Activates a filter that enables a fast change in the display for very fast load
changes. Display changes with minimal load changes (in the digit range) occur
more slowly.
Dosing Activates a filter that enables a very fast change in the display with minimal load
changes, e.g., when dosing or filling vessels.
Stability signal High accuracy
Medium
accuracy*
Fast
* Factory setting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Entris II BCA Advanced Series and is the answer not in the manual?

Sartorius Entris II BCA Advanced Series Specifications

General IconGeneral
BrandSartorius
ModelEntris II BCA Advanced Series
CategoryScales
LanguageEnglish

Related product manuals