Entris
®
II Advanced Line Operating Instructions 21
Operating Concept
4.11.4 Parameters in the “Settings / Weighing” Menu
Parameters Settings Explanation
Safety Level High* If the balance is not correctly leveled or a calibration/adjustment is not needed
or the minimum initial weight requirements according to USP are not met:
A warning message is displayed.
No data is transmitted.
The printout is blocked.
Starting and saving in applications is blocked.
Standard If the balance is not correctly leveled or a calibration/adjustment is not needed
or the minimum initial weight requirements according to USP are not met:
A warning message is displayed.
Low If the balance is not leveled: The [Start leveling] button is active.
If the balance needs to be calibrated and adjusted: The [isoCAL] button is
active.
Ambient
conditions
Very stable Sets the ambient conditions to “very stable”. Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
Very stable table near the wall
Closed and calm room
Stable* Sets the ambient conditions to “stable”. Recommended for the following work
environment:
Stable table
Slight movement in the room
Slight draft
Unstable Sets the ambient conditions to “unstable”: Activates the delayed change in
weight values with a reduced output rate. Recommended for the following work
environment:
Simple office desk
Room with moving machinery or personnel
Slight air movement
Very unstable Sets the ambient conditions to “very unstable”: Activates a significantly delayed
change in the weight values and long wait for stability with a further reduction in
the output rate.
Recommended for the following work environment:
Noticeable and slow floor vibrations
Noticeable building vibrations
Weighed goods moved
Very strong air movements
Application Weighing* Activates a filter that enables a fast change in the display for very fast load
changes. Display changes with minimal load changes (in the digit range) occur
more slowly.
Dosing Activates a filter that enables a very fast change in the display with minimal load
changes, e.g., when dosing or filling vessels.
Stability signal High accuracy
Medium
accuracy*
Fast
* Factory setting