II Advanced Line Operating Instructions 53
Technical Data
14.12.4 Models BCA12201 | BCA10201 | BCA8201
BCA12201I-1x |
BCA12201-1x
BCA10201I-1x |
BCA10201-1x
BCA8201-1x
BCA8201I-1x
Unit Value Value Value Value
Readability | Scale interval (d) mg 100 100 100 100
Maximum capacity (Max) g 12,200 10,200 8,200 8,200
Repeatability
At 5%, typical value mg 50 50 50 50
At approx. the maximum load, typical value mg 100 100 100 100
Linearity deviation
Limits ± mg 100 100 300 100
Typical value ± mg 60 60 100 60
Sensitivity drift between +10°C and +30°C ±
ppm/K
4 4 7 4
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change K 2 2 2
Time interval h 6 6 6
Only for models with approval:
Accuracy class II II II II
Type BC-AG BC-AG BC-AI BC-AG
Verification scale interval (e) mg 1,000 1,000 1,000 1,000
Minimum load (Min) mg 5,000 5,000 5,000 5,000
Minimum initial weight according to USP (United
States Pharmacopeia), Chap. 41
Optimum minimum initial weight g 82 82 82 82
Typical stabilization time 100 100 100 100
Typical measurement time s    
Typical stabilization time s    
Recommended calibration weight
External calibrated test weight g 10,000 10,000 5,000 5,000
Accuracy class in accordance with OIML R111-1 F2 F2 F2 F2