Loading...
Entris
®
II Advanced Line Operating Instructions 51
Technical Data
14.12.2 Models BCA2202 | BCA1202 | BCA822
BCA2202-1x
(x = only S, SAR, SJP, SKR)
BCA2202I-1x | BCA 2202-1x
(x = only CCN, OIN)
BCA1202-1x
(x = only S, SAR, SJP, SKR)
BCA1202I-1x | BCA 1202-1x
(x = only CCN, OIN)
BCA822-1x
(x = only S, SAR, SJP, SKR)
BCA822I-1x | BCA 822-1x (x =
only CCN, OIN)
Unit Value Value Value Value Value Value
Readability | Scale interval (d) mg 10 10 10 10 10 10
Maximum capacity (Max) g 2,200 2,200 1,200 1,200 820 820
Repeatability
At 5%, typical value mg 5 5 5 5 5 5
At approx. the maximum load, typical value mg 10 10 10 10 10 10
Linearity deviation
Limits ± mg 20 20 20 20 20 20
Typical value ± mg 6 6 6 6 6 6
Sensitivity drift between +10°C and +30°C ±
ppm/K
3.5 2 3.5 2 3.5 2
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change K 2 2 2
Time interval h 6 6 6
Only for models with approval:
Accuracy class II II II
Type BC-AE BC-AE BC-AE
Verification scale interval (e) mg 100 100 100
Minimum load (Min) mg 500 500 500
Minimum initial weight according to USP (United
States Pharmacopeia), Chap. 41
Optimum minimum initial weight g 8.2 8.2 8.2 8.2 8.2 8.2
Typical stabilization time 10 10 10 10 10 10
Typical measurement time s      
Typical stabilization time s      
Recommended calibration weight
External calibrated test weight g 2,000 2,000 1,000 1,000 500 500
Accuracy class in accordance with OIML R111-1 F1 F1 F1 F1 F2 F2