22 Entris
®
 II Advanced Line Operating Instructions
  Operating Concept
4.11.5  Parameters in the “Settings / Weighing / Zero/Tare” Menu
Parameters Settings Explanation
Zero / tare 
function
Without 
stability
The function of the [Zero] or [Tare] button is executed immediately once the 
button is pressed.
With stability* The function of the [Zero] or [Tare] button is executed if stability exists when the 
button is pressed.
Automatic 
zero
On* Activates automatic zeroing. The display is automatically set to zero in case of a 
deviation of 0 less than (X).
Off Deactivates automatic zeroing. Zeroing must be triggered with the [Zero] 
button.
Zero / tare at 
power on
On* Activates the initial taring / zeroing. The device is tared or zeroed after it is 
switched on.
Off Deactivates the initial taring / zeroing. After it is switched on, the device shows 
the same value as before it was last switched off.
SQmin** On* Activates the SQmin application.
Off Deactivates the SQmin application.
* Factory setting
4.11.6  Parameters in the “Settings / Printout” Menu
Parameters Settings Explanation
Printout Manual 
without 
stability
The print process can be started manually at any time.
Manual with 
stability*
If the [Print] button is pressed: The print command is only executed once stability 
is achieved.
Auto without 
stability
  The data is printed automatically. 
  No stability is required for printout.
  The printout can be without unit symbols.
Auto without 
stability
  The data is printed automatically. 
  No stability is required for printout.
Print after 
weight 
change
Prints after a threshold for device stability has been exceeded and the weight has 
previously been reduced to less than half of the threshold.
* Factory setting