22 Entris
®
II Advanced Line Operating Instructions
Operating Concept
4.11.5 Parameters in the “Settings / Weighing / Zero/Tare” Menu
Parameters Settings Explanation
Zero / tare
function
Without
stability
The function of the [Zero] or [Tare] button is executed immediately once the
button is pressed.
With stability* The function of the [Zero] or [Tare] button is executed if stability exists when the
button is pressed.
Automatic
zero
On* Activates automatic zeroing. The display is automatically set to zero in case of a
deviation of 0 less than (X).
Off Deactivates automatic zeroing. Zeroing must be triggered with the [Zero]
button.
Zero / tare at
power on
On* Activates the initial taring / zeroing. The device is tared or zeroed after it is
switched on.
Off Deactivates the initial taring / zeroing. After it is switched on, the device shows
the same value as before it was last switched off.
SQmin** On* Activates the SQmin application.
Off Deactivates the SQmin application.
* Factory setting
4.11.6 Parameters in the “Settings / Printout” Menu
Parameters Settings Explanation
Printout Manual
without
stability
The print process can be started manually at any time.
Manual with
stability*
If the [Print] button is pressed: The print command is only executed once stability
is achieved.
Auto without
stability
The data is printed automatically.
No stability is required for printout.
The printout can be without unit symbols.
Auto without
stability
The data is printed automatically.
No stability is required for printout.
Print after
weight
change
Prints after a threshold for device stability has been exceeded and the weight has
previously been reduced to less than half of the threshold.
* Factory setting