EasyManua.ls Logo

Sartorius Entris II BCA Advanced Series - Operation; Switching the Device On;Off (Standby); Waiting for the Warm-up Time; User Login

Sartorius Entris II BCA Advanced Series
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38 Entris
®
II Advanced Line Operating Instructions
Operation
8 Operation
8.1 Switching the Device On/Off
(Standby)
Requirements
The device is connected to the power supply.
Procedure
t
NOTICE Pointed or sharp-edged objects may
damage the operating display! Only touch the
operating display with your fingertips.
t To switch the device on: Press the [On/Off] button.
t To switch the device to standby mode: Press and
hold the [On/Off] button.
8.2 Waiting for the Warm-up Time
In order to provide accurate weighing results, the device
must have reached the necessary operating
temperature. The operating temperature is reached
after the warm-up time. The warm-up time starts when
the device is switched on.
Procedure
t Switch on the device.
t Wait until the operating temperature is reached (for
warm-up time, see Chapter “14.8 Warm-up Time”,
page 48).
The warm-up time must be observed for a
conformity-assessed device, otherwise the
weight values cannot be used.
8.3 User Login
If user profiles are created: A user login is required every
time the device is switched on.
Procedure
t Switch on the device.
t Press the user profile
selection (1).
t Select a user profile,
e.g., Administrator.
t Press the [Apply] button (2).
y If a password is assigned: The input screen for the
password opens.
t Enter the password and confirm with the [Confirm]
button.
t To log in a different user: Press the [On/Off] button.
y The device is in standby mode.
t Press the [On/Off] button.
y The device is switched on.
y User Management is displayed.
t Log in a new user.
8.4 Leveling the Device with an
Electronic Level Indicator
Unevenness at the installation site of the device may
result in incorrect weighing results. Leveling the balance
compensates for unevenness at the installation site by
twisting the front leveling feet of the balance.
Procedure
t Follow the instructions of the Leveling Wizard.
8.4.1 Level Indicator Test
The level indicator test is used to check the movement
of the level indicator. The level indicator must be
positioned at three different places during the level
indicator test. The level indicator is positioned by
rotating the front leveling feet.
Procedure
t Open the main menu.
t Press the “Leveling” button
t Press “Level test”.
t Follow the Level Indicator Wizard’s instructions.
1 2