40 Entris
®
 II Advanced Line Operating Instructions
 Operation
Procedure
 t Press the [Toggle be-
tween weight units] 
button (1).
 t Set the unit and the 
accuracy.
 t Confirm with the 
[Confirm] button.
8.7.2  Running the “Statistics” Application
The following values are saved and analyzed:
  Number of components
  Mean value
  Standard deviation
  Variation coefficient sum of all values
  Lowest value (minimum)
  Highest value (maximum)
  Spread: Difference between maximum and minimum
Requirements
  A printer or a PC is connected and configured.
  The “Statistics” application is selected.
Procedure
 t Place a sample on the weighing pan.
 t To save the weight value: Press the [Apply] button.
 y The position of the saved value is displayed, e.g., 
“Component 1”.
 y The recording of the statistics starts.
 t Remove the sample being weighed.
 t To save the next value: Place a new sample on the 
weighing pan and press the [Apply] button.
 t To display the statistical analysis: Press the [Report] 
button.
 t To print the statistical analysis: Press the [Print] 
button.
 t To exit the application and delete the saved values: 
Press the [End] button. 
1
8.7.3  Running the SQmin Minimum Initial 
Weight Application
The SQmin application compares the current weight 
value with a defined minimum initial weight. The defined 
minimum initial weight is defined in a quality assurance 
system, for example. This ensures that the weight values 
are above the defined minimum initial weights and thus 
meet the requirements of the United States 
Pharmacopeia (USP) 
SQmin is not the same as the minimum load 
“Min” in legal metrology.
Requirements
  The SQmin application must only be set up and 
  The “SQmin” application is activated.
Procedure
 t Place a sample on the weighing pan.
 y If a red SQmin value is displayed: The weight value is 
less than the defined minimum initial weight. The 
weight value is displayed in gray. The weight value 
cannot be saved and cannot be printed out in 
applications (depending on the setting in the “Safety 
level” menu).