52 Entris
®
II Advanced Line Operating Instructions
Technical Data
14.12.3 Models BCA5201 | BCA2201
BCA5201-1x (x = only
S, SAR, SJP, SKR)
BCA5201I-1x |
BCA5201-1x (x = only
CCN, OIN)
BCA2201-1x (x = only
S, SAR, SJP, SKR)
BCA2201I-1x |
BCA2201-1x (x = only
CCN, OIN)
Unit Value Value Value Value
Readability | Scale interval (d) mg 100 100 100 100
Maximum capacity (Max) g 5,200 5,200 2,200 2,200
Repeatability
At 5%, typical value mg 50 50 50 50
At approx. the maximum load, typical value mg 100 50 100 50
Linearity deviation
Limits ± mg 300 100 300 100
Typical value ± mg 100 60 100 60
Sensitivity drift between +10°C and +30°C ±
ppm/K
7 2 7 2
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change K 2 2
Time interval h 6 6
Only for models with approval:
Accuracy class II II
Type BC-AE BC-AE
Verification scale interval (e) mg 100 100
Minimum load (Min) mg 5,000 5,000
Minimum initial weight according to USP (United
States Pharmacopeia), Chap. 41
Optimum minimum initial weight g 82 82 82 82
Typical stabilization time 100 100 100 100
Typical measurement time s    
Typical stabilization time s    
Recommended calibration weight
External calibrated test weight g 5,000 5,000 2,000 2,000
Accuracy class in accordance with OIML R111-1 F2 F2 F2 F2