52 Entris
®
II Advanced Line Operating Instructions
Technical Data
14.12.3 Models BCA5201 | BCA2201
BCA5201-1x (x = only
S, SAR, SJP, SKR)
BCA5201I-1x |
BCA5201-1x (x = only
CCN, OIN)
BCA2201-1x (x = only
S, SAR, SJP, SKR)
BCA2201I-1x |
BCA2201-1x (x = only
CCN, OIN)
Unit Value Value Value Value
Readability | Scale interval (d) mg 100 100 100 100
Maximum capacity (Max) g 5,200 5,200 2,200 2,200
Repeatability
At 5%, typical value mg 50 50 50 50
At approx. the maximum load, typical value mg 100 50 100 50
Linearity deviation
Limits ± mg 300 100 300 100
Typical value ± mg 100 60 100 60
Sensitivity drift between +10°C and +30°C ±
ppm/K
7 2 7 2
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change K 2 2
Time interval h 6 6
Only for models with approval:
Accuracy class II II
Type BC-AE BC-AE
Verification scale interval (e) mg 100 100
Minimum load (Min) mg 5,000 5,000
Minimum initial weight according to USP (United
States Pharmacopeia), Chap. 41
Optimum minimum initial weight g 82 82 82 82
Typical stabilization time 100 100 100 100
Typical measurement time s    
Typical stabilization time s    
Recommended calibration weight
External calibrated test weight g 5,000 5,000 2,000 2,000
Accuracy class in accordance with OIML R111-1 F2 F2 F2 F2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Entris II BCA Advanced Series and is the answer not in the manual?

Sartorius Entris II BCA Advanced Series Specifications

General IconGeneral
BrandSartorius
ModelEntris II BCA Advanced Series
CategoryScales
LanguageEnglish

Related product manuals