50 Entris
®
 II Advanced Line Operating Instructions
  Technical Data
14.12  Metrological Data
14.12.1  Models BCA6202 | BCA4202 | BCA3202
BCA6202I-1x | 
BCA6202-1x
BCA4202I-1x | 
BCA4202-1x
BCA3202I-1x | 
BCA3202I-1x
Unit Value Value Value
Readability | Scale interval (d) mg 10 10 10
Maximum capacity (Max) g 6,200 4,200 3,200
Repeatability
At 5%, typical value mg 5 5 5
At approx. the maximum load, typical value mg 10 10 10
Linearity deviation
Limits ± mg 20 20 20
Typical value ± mg 6 6 6
Sensitivity drift between +10°C and +30°C ± 
ppm/K
2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for I-1x models): 
Temperature change K 2 2 2
Time interval h 6 6 6
Only for models with approval:
Accuracy class II II II
Type BC-AE BC-AE BC-AE
Verification scale interval (e) mg 100 100 100
Minimum load (Min) mg 500 500 500
Minimum initial weight according to USP (United 
States Pharmacopeia), Chap. 41
Optimum minimum initial weight g 8.2 8.2 8.2
Typical stabilization time 10 10 10
Typical measurement time s   
Typical stabilization time s   
Recommended calibration weight
External calibrated test weight g 5,000 2,000 2,000
Accuracy class in accordance with OIML R111-1 F1 F1 F1