Calibrating Functions
There are 3 setting programs designed to ease initial
calibration and recalibration.
(1) Pulses/Belt Circuit LB
(2) Taring TW
(3) Weight Check CW
Upon initial calibration, perform functions in the above
order of sequence.
All functions can be reached using the CALIBRATION
function and are protected from unauthorized use by
password.
During program run until acknowledgement
(1) no totalization takes place
(2) analog outputs are set to offset values
(3) no event messages are suppressed
(4) all BIC functions are cut off.
The operating sequence is the same with all
programs.
G Call function distributor.
TUf Scroll CALIBRATION FUNCTIONS
into lower display field and
acknowledge.
B7f Enter password 07734.
Display shows a setting program, e.g. TW: Tare.
f
Start setting program.
N Return to normal displays.
For approx. 2 min., Calibration menu can be
recalled without password input.
Belt Circuit LB
Call setting program LB: IMP/BELT only
(1) upon initial calibration
(2) after mounting new belt or considerably varying
belt tension
(3) after having changed parameter B 04
“Charact.Val. vs” or B 05 “Nominal Speed” (in
case of scales with and without speed
measurement).
Program acquires the number of speed transducer
pulses for one belt circuit and uses the number of
pulses as belt circuit code for the following programs:
(a) Manual or automatic zero setting
(b) Taring TW
(c) Weight Check CW
Upon initial calibration, first call setting program LB.
This also applies to scales without speed
measurement.
Prerequisites : (1) Before calling program, measure
time of one belt circuit with
maximum possible accuracy and
enter into Parameter C 02.
(2) Deselect batching mode.
(3) Ensure conveyor belt is running.
Sequence:
Gf Function distributor, calibrating
functions, password
(see “Calibrating Functions”)
TU Scroll LB: Imp/Belt
into lower display field.
f Start.
N Abort at any time.
Setting programs do not control belt drive and
prefeeder.
If any condition is not met, setting program aborts
and a corresponding message is output.
0837
VBW INTECONT PLUS
Setting Programs
BV-H2214 GB 34
Technische
Redaktion PDE-RD
Transmission to third parties and reproduction of
this documentation are not permitted. Schenck
Process GmbH reserves all rights of ownership
and copyrights.
Calibrating Functions
S5 TW. Tare

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Schenck process Intecont Plus Belt Weigher and is the answer not in the manual?

Schenck process Intecont Plus Belt Weigher Specifications

General IconGeneral
BrandSchenck process
ModelIntecont Plus Belt Weigher
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals