Program running:
Upper display : residual run time in % of total run time
Lower display : continuously meaned SET/ACT
measuring result
Program complete:
Upper display : Fictitious material amount fed over
run time *.
Lower display : mean value KOR of SET/ACT over
total run time
fN Exit program.
Since result is not
AUTOMATICALLY ACQUIRED,
both keys are equivalent.
Special Messages:
(1) START if belt is not running
(2) ABORTED (irrelevant)
* Lets you check scale zero point without check
weight.
Display format: xxxxxxx,yy kg
Evaluating Result:
Error<1% : KOR=0.99...1.01
Scale OK, no further action required.
Error<5% : KOR=0.95...1.05
Enter KOR value into Parameter D
02.
This make good sense only if
parameter has not yet considered
result of check using material.
Error>5% : KOR<0.95, or KOR > 1.05
Multi-percent deviations suggest
faulty data input (e.g. belt incination,
lever arms not exactly known), and/or
mechanical faults (misalignment,
distorsions).
Check does not consider Parameter D 02 “Range
Correction”, so that check program will indicate
same error quotient KOR after check.