7. Meter Settings [Setting] Screen
Operation
1.
Touch the item [Dark Calibration] button on page 3 of Setting.
"Perform Dark Calibration?" will be displayed.
7-4 Dark Calibration
Dark calibration is performed when 24 hours passed since last use or there is a big change in
temperature between turning power OFF and ON again.
Except the cases above, dark calibration after power ON is skipped. Accordingly, If you want to
perform dark calibration manually, do it from this screen.
Light Selection Ring Status Indicator Chart
Icon Content
If either of icons appears, set
the Light Selection Ring
to
the dark calibration position.
This icon is the dark calibration
position.
Perform dark calibration after
making sure it is set to this
position.
Dark Calibration
Conrmation Screen
Setting Screen
Dark Calibration
Conrmation Screen