Make sure that the IP address of the gateway and the network interface are not the same
Verify that the appropriate Ethernet link indicators of the PC and probe are lit
Verify that web browser proxy settings are not interfering
Verify that local firewall settings on the laptop are not interfering
Make sure that the management and data/video subnets do not overlap (even if only one is
physically connected)
Make sure the probe was rebooted to activate the new settings
Clear the browser’s cache
38 VB2xx GigE User’s Manual version 5.4