17
8014084/ZWT7/V3-1/2020-06 | SICK OPERATING INSTRUCTIONS | TRANSIC111LP
Subject to change without notice
INSTALLATION 3
3.1.3 Strong light sources near the oxygen measuring probe
The interfering effect is influenced by:
Filter used
Light incidence angle into the measuring probe
Use a stainless steel filter for ambient light, e.g., normal interior or laboratory light.
Use a PTFE filter for especially strong light sources, e.g. outdoors with direct sunlight.
3.1.4 Pressure
Observe the specifications on pressure conditions, see “Flange mounted TRANSIC111LP-
A/-D/-F/-G/-H/-I/-J/-K”, page 14, see “TRANSIC111LP-C/-E with sample gas cell and wall
bracket”, page 14, see “TRANSIC111LP-B for ambient gas measurement and wall
bracket”, page 15 and see “Dimensions and mechanics”, page 102.
NOTICE: Strong light sources interfere with TRANSIC111LP operation
Prevent strong light sources reaching the measuring probe.