101
8014084/ZWT7/V3-1/2020-06 | SICK OPERATING INSTRUCTIONS | TRANSIC111LP
Subject to change without notice
SPECIFICATIONS 10
10.2.2 Measured value recording
10.2.3 Ambient conditions
Measuring ranges (scalable) 0 … 25% O
2
or 0 … 100% O
2
Precision ±0.2% O
2
Temperature dependence in T-range ±2% of measured value, max. dT/dt 1 °C/
min
Stability Zero point drift ±0.1% O
2
/year
Span gas drift ±0.8% of measured value /
year
Measurement response time (T
63
/T
90
) in still air
–without filter
– with stainless steel mesh
– with stainless steel mesh and PTFE
10 s / 20 s
10 s / 25 s
30 s / 70 s
Operating pressure range 0.8 … 1.4 bar
a
Start time 2.5 min
Warming up time (according to specification) 3 min
Display 7-segment LCD
LED Two-colored: Red/green
Operating temperature range
– for probe (installed in process)
– for electronics (housing)
– for measuring device (ambient air measurement)
–for cable
−20 … +60 °C
−20 … +60 °C
−20 … +60 °C
−20 … +60 °C
Storage temperature range −40 … +80 °C
Operating pressure range 0.8 … 1.4 bar
a
Air humidity 100% r.h. non-condensing
Altitude Up to 2000 above sea level
Electrical compliance In accordance with EN61010-1
Safety information Laser product of protection class 1;
information on eye-safe use of the device see
“Assembly and disassembly”, page 84.
WARNING: TRANSIC111LP is not suitable for use in a potentially explosive
atmosphere
The TRANSIC111LP is NOT designed or approved for use in potentially explosive
atmospheres. Use of TRANSIC111LP in potentially explosive atmospheres is not
permitted.