4 OPERATION
4.7 Output of measuring results
4.7.1 Start continuous output (command R)
Starts RUN mode. Outputs values defined with the command FORM (see “Format measur-
ing results (command FORM)”, page 45). Command INTV (see “Display/set continuous out-
put interval (command INTV)”, page 42) defines the output interval. Command S (see “Stop
continuous output (command S)”, page 42) stops RUN mode.
Syntax:
R<cr>
Example:
4.7.2 Stop continuous output (command S)
Stops RUN mode and switches the serial output to STOP.
Syntax:
S<cr>
Example:
4.7.3 Display/set continuous output interval (command INTV)
Sets the frequency for measured value output in RUN mode (see “Start continuous output
(command R)”, page 42).
Syntax:
INTV [Value] [Unit]<cr>
Example:
4.7.4 Send measuring results (command SEND)
Outputs the last results (in accordance with FORM, see “Format measuring results (com-
mand FORM)”, page 45) in STOP mode. The command can be used with an address in
POLL mode.
Syntax:
SEND [Address]<cr>
SEND [Formatting string]<cr
Example:
>r
Oxygen = 21.0
Oxygen = 21.0
Oxygen = 21.0
>S
>
Value = Time interval in which results are output (0 ... 255)
Unit = Interval time unit, S for seconds, MIN for minutes or H for hours
>intv
INTERVAL : 1 ? 5
UNIT S ? min
Address = Device address
Formatting string = Character string specifying the output format for the measuring results
>send 20.9 20.8 24.5