EasyManua.ls Logo

SICK TRANSIC111LP - Analog Output Display and Settings (AOU); Content in Process Gas: Settings (CO2); One-Point Calibration (CAL1); Resetting to the Factory Calibration (FAC)

SICK TRANSIC111LP
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
8014084/ZWT7/V3-1/2020-06 | SICKOPERATING INSTRUCTIONS | TRANSIC111LP
Subject to change without notice
4 OPERATION
4.5.3 C0
2
content in process gas: Settings (CO2)
1 Enter the average CO
2
value in the sample gas.
Adjustable range: 0 ... 100% vol.
Fig. 25: CO
2
sample gas setting
4.5.4 One-point calibration (CAL1)
Figure can be found in Section Adjustment, see “One-point adjustment using the keypad”,
page 73.
4.5.5 Two-point calibration (CAL2)
Figure can be found in Section Adjustment, see “Two-point adjustment using the keypad”,
page 76.
4.5.6 Analog output display and settings (AOU)
1 Press Ent to display the current output value on the analog output.
2 To set fixed output values for the analog output (0, 4, 12, 20 mA), press Ent and use Up
and
Dn
to select the analog output value.
Fig. 26: Displaying and setting the analog output value.
4.5.7 Resetting to the factory calibration (FAC)
The adjustment is reset to factory setting.
(Gain value: 1, offset value: 0).
Fig. 27: Resetting to the factory setting for oxygen measurement
<
(0 ... 100 vol%)
<
<
<
<
<
Back
Back
k
Ent
E
n
t
Up
Up
Dn
D
n
Ent
En
t
Dn
Back
Back
k
Up
Back
B
ac
k
Ent
E
n
t
Ent
E
n
t
Back
Back
k
<
<
<
Ent
E
n
t

Table of Contents

Related product manuals