7 MAINTENANCE
7 Rinse the optical components with distilled water after cleaning.
8 Dry with compressed air (instrument air or better quality).
9 After cleaning, the surface should appear clean without oil stains, contamination or dust.
Refit the filters after cleaning.
NOTICE: Damage to the lens and mirror through mechanical cleaning.
During the cleaning process described above, never attempt to clean the optical
components by rubbing (e.g., with cotton swabs or cleaning a cloth).
Cleaning the TRANSIC111LP optics is easiest with an optics cleaning set from SICK.
Part No., see “Accessories”, page 90.