8014084/ZWT7/V3-1/2020-06 | SICK OPERATING INSTRUCTIONS | TRANSIC111LP
Subject to change without notice
TROUBLESHOOTING 8
36 NONFATAL
(non-serious)
ANALOG OUTPUT LOAD TOO
HIGH
Analog output load too high.
Check specifications for voltage supply unit and
cables. (See Technical Data, Page 102).
37 NONFATAL
(non-serious)
NO MEASUREMENT RESULTS No measuring results (results from other errors)
38 NONFATAL
(non-serious)
ANALOG OUTPUT RANGE Oxygen concentration value measured outside set
output range.
Adjust the output range settings when necessary.
51 WARNING SIGNAL QUITE LOW Transmission (SIL) <20%
Maintenance request for optical components, see
“Cleaning the optical components”, page 85.
52 WARNING EEPROM LOG&STATS COR-
RUPTED
Non-critical hardware error: EEPROM protocol and
statistics erroneous.
(Contact SICK Customer Service)
53 WARNING WATCHDOG RESET OCCURRED Reset through software error.
Error
No.
Error category Error text Cause
Table 13: Error Table