Parameter Default val‐
ue
Value range Access
level
Description
109 Actual flow velocity for path 2 0 Read The actual flow velocity for path 2 in m/s
(SI-unit).
Valid for 2-path solutions only
110 Path 1 gain step 15 1 to 15 Read Amplifier setting when measuring path 1
111 Path 2 gain step 15 1 to 15 Read Amplifier setting when measuring path 2.
Valid for 2-path solutions only
Pulse A B
430 Output A enable Product de‐
pendent
No, Yes R/W Select "Yes" to enable output A
431 Pulse A direction Forward Forward, Re‐
verse, Forward
net, Reverse net
R/W Pulse output calculation on pure forward
flow or pure reverse flow - or net flow for
forward and reverse pulse flow.
Valid if pulse output A is enabled
432 Amount per pulse A Product de‐
pendent
0.000001 to
1000000000
R/W Volume per pulse and selected unit
Valid if pulse output A is enabled.
433 Pulse width for pulse A Product de‐
pendent
5, 10, 20, 50,
100, 200, 500 ms
R/W Pulse length when the pulse is active.
Valid if pulse output A is enabled
440 Output B enable Product de‐
pendent
No, Yes R/W Select "Yes" to enable output B
441 Pulse B function Product de‐
pendent
Pulse, Alarm,
Call up
R/W Configuration of output B as pulse - alarm
or call up function.
Valid if pulse output B is enabled
442 Pulse B direction Reverse Forward, Re‐
verse, Forward
net, Reverse net.
R/W Pulse output calculation on pure forward
flow or pure reverse flow - or net flow for
forward and reverse pulse flow.
Valid if pulse output B is enabled and
pulse function selected
443 Amount per pulse B Product de‐
pendent
0.000001 to
1000000000
R/W Volume per pulse.
Valid if pulse output B is enabled and
pulse function selected
444 Pulse width for pulse B Product de‐
pendent
5, 10, 20, 50,
100, 200, 500 ms
R/W Pulse length when the pulse is active.
Valid if pulse output B is enabled and
pulse function selected
Communication
For editing the communication settings, use PDM device menu "Communication Setting"
591 Device Communication Ad‐
dress
1 1 to 247 Read Meter has default address value 1 with
selectable address up to 247
592 Baud rate 4 0 to 5 Read Communication port speed
593 Parity 0 0 to 3 Read Communication port parity
594 Interframe space 35 35 to 255 Read Minimum space between two messages
(bytes x 10)
595 Response delay 5 1 to 50 Read Minimum time from receiving a request to
its response
Data logger
600 Log interval Monthly Daily, Weekly,
Monthly
R/W Log interval
Parameter lists
A.2 Output