EasyManua.ls Logo

SIGLENT SDS1000CML+ - Function Inspection

SIGLENT SDS1000CML+
159 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SIGLENT
SDS1000CML+/SDS1000DL+ Quick Start 7
Function Inspection
1. Press “Default Setup” to restore the oscilloscope to its
default settings.
2. Connect the ground alligator clip of the probe to the Ground
Terminal on the front panel.
3. Use the probe to connect the CH1 Input Terminal and the
Compensation Signal Output Terminal on the front panel.
4. Press “Auto”.
5. Observe waveforms on the screen. In normal condition, the
display should be several square waveforms as is shown below:
6. Test channel 2 in the same method. If the square waveforms do
not display like figure above, please perform “Probe
Compensation”.
Note: To avoid electric shock when using the probe, please firstly
make sure that the insulated wire of the probe is in good
condition, and do not touch the metallic part of the probe
when it is connected to a high voltage.

Table of Contents

Related product manuals