SIGLENT
8 SDS1000CML+/SDS1000DL+ Quick Start
Probe Compensation
You should properly compensate the probe at first use of it.
Non-compensated or inadequate compensated probe may cause
inaccurate measurement. The following steps are about probe
compensation:
1. Perform step 1, 2 , 3 and 4 of “Function Inspection”.
2. Check the displayed waveforms and compare them with the
following figure.
Under Compensated Over
Compensated Correctly Compensated
3. Use a nonmetallic driver to adjust the low-frequency
compensation adjustment hole on the probe until the waveform
changes to be correct as the figure above.