SDS1000CML+/SDS1000DL+ User Manual 89
Picture 2.13-7
2.13.3 Self Calibration
Self Calibration procedure can optimize the signal path at the most measurement
precision. You can run this procedure at any time. If the operating temperature
changes by more than 5° C or the units run more than thirty minutes, you should
do the self calibration.
When you run the self Calibration, you should cut off all the probes and lead.
Then press the “Utility” button to choose the “Do self cal” to show the self
Calibration Menu, and execute self calibration program according to the
prompting on the screen.
Picture 2.13-8