64 57 316 D3543
20 D3543.076.01.04.02 06.2015
5Important information Sirona Dental Systems GmbH
5.2Installing the treatment center software Service Manual INTEGO / INTEGO pro
Software installation via SD card on EasyTouch
For details of how to use the DUMC to prepare the SD card for the
software update process, please refer to the "Dental Unit Management
Console Service Manual" (DUMC) with order number 6194497.
The treatment center is switched off.
The connection box cover on the patient chair has been removed
(see Removing the connection box cover [ 97]).
The SD card has been prepared for the firmware update process
using the DUMC.
1. Insert the SD card in the SD card slot on the DSA board.
2. Switch the treatment center on.
Downloading the update files has started.
The updates of the individual boards are performed
consecutively.