Sirona Dental Systems GmbH 5Important information
Service Manual INTEGO / INTEGO pro 5.2Installing the treatment center software
båÖäáëÜ
In case of an error
The update for the separate CAN nodes (DSA) could not be completed
successfully.
The treatment station software no. in the download request is invalid. The
download manager performs a reset.
A serious error has occurred during the download procedure. The update
process is terminated.
If the download for a specific CAN node (board) fails, the error displayed
in the margin column is shown. The number stands for the ID of the board
(in this example, board NAA) where the download failed.
Board IDs:
dL.Er.1: DSA
dL.Er.2: DSL
dL.Er.3: DAB
dL.Er.4: DAP
dL.Er.5: NAA
dL.Er.6: DWE
dL.Er.7: NLV
dL.Er.8: DAK
No valid associated application could be found. Unable to start the
download manager
A software inconsistency between a CAN board and the complete
software of the treatment center has occurred.
Board IDs:
Err.1: DSA
Err.2: DSL
Err.3: DAB
Err.4: DAP
Err.5: NAA
Err.6: DWE
Err.7: NLV
Err.8: DAK