6Overview of modules and boards Sirona Dental Systems GmbH
6.3Dentist element Service Manual INTEGO / INTEGO pro
Function of the board
Communication
Via CAN with the DSA
Via CAN with the DAK
Via a serial interface with the DAT
Via CAN with the DAP
Via a serial interface with the DAI
Supply:
Of the DAK and DAT with 5 V
Of the DAP with 5 V for the electronics circuit of the DAP
Activation and evaluation of the instrument holders
Activation of the solenoid valves for the instruments (MV block),
spray, chip blower, and driving air
Control or evaluation of signals (F_CTRL, Chair_Stop)
Optional
Activation of the solenoid valves for the purge function and water for
the ultrasonic handpiece
Activation of the X-ray viewer