www.spacelabshealthcare.com
2-5
91390 QUBE
S ETUP
Remote Nurse Alert Connector -
14 Pin for option A
The Remote Nurse Alert connector is primarily designed to provide
a contact closure to the outside world to indicate an alarm
condition. The nurse alert output provides +12V, return and three
sets of relay contacts. There is one set of relay contacts for each of
the three alarm levels. Each set of relay contacts consists of a
common normally open (NO) and normally closed (NC). Each
contact is able to carry at least 150 mA continuous.
The pinout for this connector is as follows:
Warning:
Data interface connectors on this device are Ground (Earth) referenced!
- Only connect this device to other medical equipment suitable for
use in the Patient Vicinity.
- Check Ground (Earth) Leakage after connecting data interface
cables.
Table 2-3 Connector Pinouts
Pin
Alarm Circuit Meaning Image
1 Alarm 0 (high priority) Common
2 Normally Closed
3Normally Open
4GND
5 Alarm 1 (medium priority) Normally Closed
6Normally Open
7 Common
8GND
9 +12 V, 140 μA
10 GND
11 GND
12 Alarm 2 (low priority) Normally Open
13 Common
14 Normally Closed