2-4
764333-098
Table 2 - 2: System Options
Description Default Range Units Description
Number of chemical pumps Without pH neutralizer:
22
With pH neutralizer:
32-3
pumps
pumps
Alkaline and acid detergent are always
present, third pump is acid neutralizer.
Cycle Count 00000001 [00000000 - 99999999] N/A —
Reservice in XXXX Cycles 9999 [0001 - 9999] N/A Number of cycles before next mainte-
nance warning.
Warranty Expiration YY/MM/DD N/A This date is used to give warnings before
warranty ends.
Sump Filling Alarm 300 [60 - 6000] Sec. Maximum time to fill sump with water.
Sump Draining Alarm 60 [01 - 6000] Sec. Maximum time to drain sump water.
Sump Heating Alarm 600 [60 - 3000] Sec. Maximum time to heat sump water.
Cycle Complete Notification 60 [01:00 -6000] Sec. Maximum time buzzer is activated for the
end of cycle notification.
Detergent Monitoring Time 180 [00:01 -6000] Sec. Maximum time allowed to reach conduc-
tivity setpoint before an alarm.
pH Neutralization Limits Low = 06 [01 - 14]
Hi = 09 [06 - 09]
pH Maximum/minimum limits for pH neu-
tralization.
pH Neutralization Mix Time 45 [1 -99] Sec. Time pH must be within range before
draining.
Alkaline Neutralizer
Inject = XX Wait = XX
Inject = 05 [02 - 99]
Wait = 45 [30 - 99]
Sec.
When injecting neutralizer, pump is
pulsed ON and OFF. Inject time is the
time pump is ON. Wait time is the time
pump is OFF.
Acid Neutralizer
Inject = XX Wait = XX
Inject = 10 [02 - 99]
Wait = 45 [30 - 99]
Sec.
ph Neutralization Number of
Tries
09 [03 - 15] N/A Maximum number of injections allowed
before giving an alarm.
High pH Containment Inject-
ing Time
02 [01 - 98] Sec.
Shorter injection time used when neutral-
izing treatment without detergent or after
pH overshoot.
Low pH Containment Inject-
ing Time
06 [01 - 98] Sec.