2-1
Installation Verification Operator Manual 920509-645
2.1 Technical
Specifications
Complete uncrating and installation instructions, as well as an equip-
ment drawing have been furnished. If any of these documents are
missing or misplaced, contact STERIS giving serial, equipment and
model numbers of the unit. Replacement copies will be sent to you
promptly.
These specifications are intended to describe technical information
given on nameplate of your washer and to state other relevant infor-
mation. Check equipment drawing or identification nameplate for
proper voltage and amperage.
2.1.1 Voltage, Amperage and
Power Consumption
Basil
®
4700 Cage and Rack Washer operates on:
208 V, 60 Hz, 3-phase, 25 Amps;
480 V, 60 Hz, 3-phase, 15 Amps.
A protective ground is required (Class 1 Equipment).
Maximum currents and power consumptions are indicated on
nameplate.
Main supply voltage not exceeding ± 10% of nominal voltage.
Installation category: Overvoltage Category II.
Always follow local electrical installation codes.
Refer to Uncrating/Installation Instructions (P920509-642) for proper
connection.
IMPORTANT: A listing of the Safety Precautions to be observed when operating this Cage and Rack
Washer can be found in S
ECTION 1. Do not operate equipment until you have become familiar with this
information.
INSTALLATION VERIFICATION