6-3
764333-098
and header spray jets as the carriage assembly
moves back and forth.
c. Press DRAIN touch pad. Once Drain function
automatically stops, open chamber door. No
water should remain in chamber sump.
d. Exit manual control mode by pressing X.
8. Select any pre-programmed cycle and press
START. Run one complete cycle and check
mechanical operation of washer.
9. Run several cycles, testing all available options.
Make sure washer functions properly and is leak-
free.
10. On distribution terminal, verify voltage and amper-
age of water pump are within specifications:
• VOLTAGE AMPERAGE MAX.
208 V (187 to 229 V) 5.8 A
480 V (432 to 528 V) 2.5 A
380-400-415 V (342 to 457 V) 3.0 A
11. Start a cycle in Automatic Mode and record pres-
sure of each supply at appropriate times during
cycle testing.
• Dynamic pressures should be:
Hot water: 15 to 50 psig (103 to 345 kPa);
Cold water: 30 to 50 psig (207 to 345 kPa);
Steam: 30 to 80 psig (207 to 550 kPa).
Figure 6-1. Manual Functions
Press MANUAL
Touch Pad
Press Touch Pad
Corresponding to
Desired Manual
Function
Press X to Exit
Manual Mode