2-3
Installation Verification Operator Manual 920509-645
Building hot water line supplies water to washer at pressure and
temperature specified on equipment drawing.
Building cold water line supplies water to washer at pressure
specified on equipment drawing.
Electrical supply for washer is as specified on equipment draw-
ing.
Condensate returns are sized as specified on equipment draw-
ing.
Vent connections are sized as specified on equipment drawing.
Recirculation pump pressure is within 25 to 60 psig.
Recirculation pump motor rotating in direction shown by arrow.
Self-cleaning screen assembly functioning properly.
Carriage drive motor rotating in direction shown by arrow.
Carriage drive system functioning properly.
Carriage drive motor amperage within rating indicated on the
motor.
Optional exhaust fan rotating in direction shown by arrow.
All piping is leak-free.
Chamber sump steam coil functioning properly.
Door safety switch(es) functioning properly.
Cabinet joints are completely sealed, no leaks (for verification,
run machine for 1/2 hour).
Door(s) easily opens from inside of chamber.
Safety cables immediately stop washer operation when pulled.
Each gear box plastic cap removed and replaced with air vent
provided.
Floor surrounding unit has non-slip surface.
IMPORTANT: After a few weeks of operation, inspect unit for leaks.
Retighten all clamps and connections.