7-6
920509-645 Operator Manual Troubleshooting
24. Jets not spraying. 1. Temperature in excess of 195°F (91°C) – Lower set-point tempera-
ture.
2. Self-cleaning screen blocked – remove, clean and reinstall prop-
erly.
3. Spray jets clogged – poke debris into jets, remove plugs located at
bottom of spray header, run spray manually and reinstall plugs.
4. Plugs in spray header missing – Reinstall plugs.
5. Jet ball valve is out of sequence – call STERIS.*
25. Sump not draining properly. 1. Self-cleaning screen blocked - Remove, clean and reinstall prop-
erly.
2. Drain valve clogged - call STERIS.*
3. Drain time set too short - call STERIS.*
4. Facility drain plugged - flush out drain line.
26. Unit not cleaning properly. 1. Temperature in excess of 195°F (91°C) – lower set-point tempera-
ture.
2. Self-cleaning screen blocked – remove, clean and reinstall prop-
erly.
3. Drain valve clogged – call STERIS.
4. Spray jets clogged – Poke debris into jets, remove plugs located at
bottom of spray header, run spray manually and reinstall plugs.
5. Spray header jammed – Verify inside chamber for obstruction.
6. Drain time set too short – call STERIS.*
7. Too much soap – rinse with cold water and decrease usage.
8. Plugs in spray header missing – reinstall plugs.
* Service Charges may be incurred. Consult your warranty for details.
Table 7-1. Troubleshooting Chart (Cont’d)