1. Mount the yellow cap (supplied) on the water inlet for
the main water supply.
2. Remove the quick coupling from one end of the hose
delivered with the pump.
3. Slide the hose clamp onto the hose and connect to
the water inlet for recirculation water on the back of
the machine. Tighten the hose clamp.
4. Connect the quick coupling at the other end of the
inlet hose directly to the pump outlet of the cooling
unit.
A Pump outlet
5.7.2 Connecting the recirculation cooling unit to the water outlet
1. Connect the water outlet hose to the water outlet
pipe. Use a hose clamp to secure the hose.
2. Lead the other end of the hose into the mounting hole
in the bracket on top of the static filter unit.
3. Make sure that the hose slopes downward towards
the waste water drain throughout its entire length.If
needed, shorten the hose.
A Static filter unit
5.7.3 Connecting the communication cable
• Connect the communication cable from the control box of the recirculation cooling unit to the
socket on the back of the machine.
5.8 Installing the preparation disc
Note
Make sure that the cavity on the underside of the preparation disc and the cone on
the machine are clean.
Make sure that the bowl liner is clean and that the drain is positioned correctly.
Procedure
1. Place the preparation disc carefully on the driving pin.
2. Rotate it slowly until it is safely engaged.
5.8.1 Preparation disc types
The machine can be used with the following disc types:
5 Installation
LaboPol-30 23