Struers LaboPol-30
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WARNING
Safety critical components must be replaced after a maximum lifetime of 20 years.
Contact Struers Service.
Note
Testing should always be performed by a qualified technician (electromechanical,
electronic, mechanical, pneumatic, etc.).
12.5.2 Emergency stop
Test 1
1. Press the Start button. The machine starts operating.
2. Press the emergency stop.
3. If operation does not stop, press the Stop button.
4. Contact Struers Service.
Test 2
1. Press the emergency stop.
2. Press the Start button.
3. If the machine starts, press the Stop button.
4. Contact Struers Service.
12 Maintenance and service