Adjusting the horizontal position of the specimen mover plate
1. Moves the specimen mover head to the right.
• Place the specimen mover plate in a position that
allows the specimen to run 3 – 4 mm over the edge
of the preparation disc.
A Disc edge
B Specimen mover plate
Complete the adjustment
1. Tighten the 2 fastening screws firmly. The specimen mover will now remain in position.
2. Cover the holes with the two caps.
Allen key and cover caps are included in the package.
9.2 Operate the device
9.2.1 Control panel functions
CAUTION
Keep clear of rotating parts during operation.