Procedure
1. From the Main menu select Maintenance > Cleaning of tubes.
2. Select the tubes you wish to clean.
Bottle No.: Identification of the bottle in the doser.
Status: Clean or Used.
Selected: No or Yes.
3. Press F1 to start the cleaning process.
4. Follow the on-screen instructions.
11.2.3 Change the tubes
When you use alcohol-based lubricants, the Novoprene tubes mounted in the pumps will harden over
time. Silicone has a better resistance against alcohol.
You can replace the tubes with the set of silicone tubes supplied with the unit.
Procedure
1. Remove the back plate.
2. Remove the tube from the pump unit:
The white connector must remain on the
tube connected to LaboForce-100.
3. Press the two tabs at the base of the
pump and remove the pump from the
axle.
A Tabs
4. Remove the three rollers.
84 LaboPol-30
11 LaboDoser-100