EasyManuals Logo

Stryker 1488 User Manual

Stryker 1488
832 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
EN-40
Guidance and Manufacturers Declaration: Electromagnetic Immunity
1488 HD Camera is intended for use in the electromagnetic environment specied below.
The customer or the user of 1488 HD Camera should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic
Environment:
Guidance
Electrostatic Discharge
(ESD) IEC61000-4-2
±6kV contact
±8kV air
±2,4,6kV contact
±2,4,8kV air
Floors should be
wood, concrete, or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast transient/
burst IEC61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
±2kV line to ground
±1kV line to line
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC61000-4-5
±1kV differential mode
±2kV common mode
±0.5, 1kV differential
mode
±0.5, 1, 2kV common
mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC61000-4-11
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of 1488
HD Camera requires
continued operation
during power mains
interruptions, it is
recommended that
1488 HD Camera
be powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60Hz) magnetic eld
IEC 61000-4-8
3 A/m N/A Power-frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE: Ut is the AC mains voltage prior to application of the test level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker 1488 and is the answer not in the manual?

Stryker 1488 Specifications

General IconGeneral
Resolution1920 x 1080
Sensor TypeCMOS
White BalanceAutomatic/Manual
OutputHDMI, DVI, HD-SDI
Camera Head TechnologyThree-chip

Related product manuals