16 72-1048E MM FL23SE REV A www.stryker.com
Return To Table of Contents
Problem/Failure Recommended Action
The positioning mechanisms of the Fowler, and/or
the Knee Gatch, and/or the bed height operate with
difficulty.
Check the four actuator screws to verify their lubrication •
condition and apply grease if needed (see pg. 20).
No Auto Contour (optional) motion. Is the Knee Gatch lockout activated (LED off)? If so, •
deactivate it (LED on).
Check and replace if needed the two Auto Contour limit •
switches (see pg. 52).
Improper operation of the CPR emergency release
(optional): Knee Gatch does not lower and/or the Fowler
actuator does not reset.
Check and replace if needed the two CPR mechanism •
limit switches (see pg. 54).
Troubleshooting Guide